5.

Operation

Access the RUN OPTIONS function to display
the options selected for the current program.

LEVEL CHECK enabled:
The lower level sensor in the retort is acti-
vated.

RECIRCULATION enabled:
During each processing step, the reagent
is pumped periodically from the bottle to the
retort and back for continuous and efficient
mixing of fluid. The first pump step is carried
out after 16 minutes, all subsequent pump
steps in 12-minute intervals.

PRETEST enabled:
After starting the program a test run is per-
formed: the first reagent is pumped into the
retort and back into the bottle.

STIRRER enabled:
Magnetic stirrer is on.

PRINTER enabled:
At the end of the program run, a full run log
will be printed out.

Important!
The options selected will be applied to ALL programs!
All run options can be enabled / disabled during the program run if necessary.

The printer option will be displayed only if a printer has been configured for the instrument.