Route]:bias:port – Agilent Technologies Agilent E5250A User Manual

Page 195

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

7-27

Command Reference

[:ROUTe]:BIAS:PORT

[:ROUTe]:BIAS:PORT

NOTE

This command applies only to the E5252A card.

For the E5255A card, the input Bias Port cannot be selected, and is always the
built-in Bias Port.

The Bias Port and Couple Port cannot be set to same port number.

This command selects which input port is the Bias Port on the specified card. For
each card, you can specify the same or different Bias Port. For details, refer to "Bias
Mode" in Chapter 5.

At *RST, Bias Port is set to 10 for the E5252A.

Syntax

[:ROUTe]:BIAS:PORT card_number,bias_port

Example

OUTPUT @Hp5250;":ROUT:BIAS:PORT ALL,6"

Parameter Explanation

card_number

card number:

For Auto Config mode: 0 or ALL

For Normal Config mode: 1, 2, 3, 4, or ALL

bias_port

Input port number: 1 to 10.