E5250a card/mode/port related errors – Agilent Technologies Agilent E5250A User Manual

Page 316

Advertising
background image

11-12

Agilent E5250A User’s Guide, Edition 11

Error Messages
E5250A Specific Error Messages

E5250A Card/Mode/Port Related Errors

Error

Number

Error Message and Description

3000

Card0 initialization fail

The E5250A may be defective. Contact your nearest Agilent
Technologies service center.

3001

Card1 initialization fail

The E5250A or card installed in slot 1 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.

3002

Card2 initialization fail

The E5250A or card installed in slot 2 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.

3003

Card3 initialization fail

The E5250A or card installed in slot 3 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.

3004

Card4 initialization fail

The E5250A or card installed in slot 4 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.

3011

Bad couple port number

For the Input Port No. for Couple Port, only 1, 3, 5, 7, or 9 are
allowed.

3012

Bad bias port number

For Input Port No. for Bias Port, only 1 to 10 are allowed.