Route]:function – Agilent Technologies Agilent E5250A User Manual

Page 207

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

7-39

Command Reference

[:ROUTe]:FUNCtion

[:ROUTe]:FUNCtion

This command sets the channel configuration, which determines how you specify
the channel_list or card_number in other commands. For details, refer to "Channel
List and Channel Configuration" in Chapter 5.

When configuration is changed by :ROUT:FUNC, all channels are opened
(disconnected) and become same status as after SYSTem:CPON ALL command
execution.

At *RST, this parameter is set to NCONfig.

Syntax

[:ROUTe]:FUNCtion channel_config

Example

OUTPUT @Hp5250;":ROUT:FUNC ACON"

[:ROUTe]:FUNCtion?

This query command returns the present channel configuration.

For details, refer to "Channel List and Channel Configuration" in Chapter 5.

Syntax

[:ROUTe]:FUNCtion?

Query response

ACON or NCON <newline><^END>

Example

OUTPUT @Hp5250;":ROUT:FUNC ACON"

OUTPUT @Hp5250;":ROUT:FUNC?"

ENTER @Hp5250;A$
In above example, A$ will be as follows:

ACON

Parameter Explanation

channel_config

ACONfig: Auto Config Mode

NCONfig: Normal Config Mode

Advertising