Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual

Page 11

Advertising
background image

7

Cycle Selection

Table 1: StatSpin MP Cycle Settings. The operator should experiment with different settings to achieve optimum
performance for specific applications. The following are general guidelines:

Setting

RPM/RCF

Time

Rotor

Urine

9,800/3,900

45 seconds

RT12

Hematocrit

16,000/13,700

120 seconds

RH12

Tube Rotor

15, 800/12,000

30 seconds

RT12/RM02

PlasmaRotor

®

20,000/11,200

30 seconds

RD01

LipoClear

®

15,800/12,000

95 seconds

RT12

Coag Plasma

20,000/11,200

120 seconds

RD01

Instructions for Use

1. Lift cover and install rotor.
2. Close and latch cover.
3. Choose desired cycle by depressing the “SET” button until the appropriate LED is illuminated.
4. Press

start button.

5. Upon completion of the cycle, the rotor decelerates to a complete stop in 10 seconds and the latch

interlock automatically unlocks.

6. Squeeze the black latch pieces together to open cover.

Advertising