Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual

Page 7

Advertising
background image

3

Symbols and Definitions (cont.)

IVD

For in vitro
diagnostic use

Clarifies for use as in vitro diagnostic only

Non sterile

Indicates non-sterile product

S N

Serial Number

Indicates instrument serial number code

Consult
Instructions

Consult instruction manual or insert sheet for further
explanation

Biological Risk

Universal precautions should be followed on all
specimens

Error Indicators

Error codes for the StatSpin MP are specified by a combination of the Error/Service LED flashing or
continuous and any combination of the cycle LEDs on the front panel.

Error/Service =































Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of 18,000, which
is the useful life of the drive system. Drive mechanism needs replacing. Contact an authorized service center.

Error/Service

light flashing

or continuous

Rotor failed
to reach rpm
within 30 s

Cover opened
or cycle
changed during
operation

Insufficient
power to
maintain
rpm

Rotor spun
backward
direction during
reverse braking