ITC AVOXimeter 4000 Operator’s Manual

Specifying a Different Value for Hüfner’s Number

f oxygen that can be carried by one gram of hemoglobin) is used

t of a sample (see page 1). Although Hüfner’s number is generally

assumed to be 1.39 mL/gm, facilities may use a different value and may wish to set the ITC
AVOXimeter 4000

to match other instruments in the facility.

1.

Display the “Calibration” menu (a submenu of

Hüfner’s number (the volume o
to calculate the oxygen conten

, see page 7).

2.

Press

followed by

to display the screen for accep g

tin or changing the

value used for Hüfner’s number:

followed by

3.

Press

to display the screen for entering a different value:

Note:

A default value of 1.39 is entered at the factory. Values ranging from
1.30 to 1.39 can be entered.

4.

Press the number key (e.g.,

) corresponding to the new value:

5.

Check that the new Hüfner’s numb r is correctly displayed, then press

e

followed by

. The “Calibration” menu is again displayed.

6.

Press

to return to the previous menu, if desired.

28