ITC AVOXimeter 4000 Operator’s Manual

13.

Secure the new battery in the battery well, using the new white plastic ties.

Note:

When securing the battery to the battery well, be sure that the plastic ties
are routed underneath the battery well. Cut the excess length from the
plastic ties using the diagonal cutting pliers.

14.

Thread the blue and white battery cable of the new battery under the main circuit
board.

15.

Re-connect the cables, replace the upper cover of the ITC AVOXimeter, connect
the AC adapter, and turn on the instrument.as outlined in Steps 17 through 25 of
the Cleaning the Optical Detector procedure on page 49. The self-test should run
and the “Ready – Insert Cuvette” screen should be displayed.

Note:

Be sure to connect the blue and white battery cable last.

The new battery is not charged. To ensure adequate charge, leave the
instrument connected to the AC Adapter for a minimum of eight hours.
Tests can be run while the instrument is charging.

16.

Reset the time and date (see page 16).