ITC AVOXimeter 4000 Operator’s Manual

4.

The screen for specifying the sample type is displayed:

5.

Press

followed by

. The screen for specifying QC type is displayed:

6.

Press

followed by

. The screen for specifying filter type is displayed:

7.

Press the number key corresponding to the filter being run, followed by

.

Press

followed by

to confirm your selection. A confirmation screen is

displayed, and test re

sults are displayed within ten seconds.

8.

Record the THb, %O

2

Hb, %COHb, and %MetHb results, or print the results of the

test (see page 35).

9.

Press any key to clear the results, then repeat steps 2 through 8 using the orange

ow:

filter.

10.

Verify that the results for each filter are within the expected ranges listed bel

Expected Range

Optical Filter

tHb (g/dL)

%O

2

Hb

%COHb

%MetHb

Yellow

7.8 to 8.2

93.7 to 96.3

0.6 to 3.4

-0.4 to 2.4

Orange

16.7 to 17.3

37.8 to 40.2

20.0 to 23.0

0.2 to 1.8

Note:

The Expected Range is also shown on a sticker on the side of each filter.

11.

If desired, plot the optical filter results on daily logs such as the ones shown in the
Quality Control Logs section on page 57.