ITC AVOXimeter 4000 Operator’s Manual

8.

Press th
by

e number key corresponding to the liquid control level being run, followed

. A menu for selection of the liquid control lot number is displayed:

9.

Select the previously entered (see page 38) lot number and press

. A

cuvette lot number screen is displayed:

If the previously entered cuvette lot number is to be used, press

followed by

. A confirmation screen is displayed.

If a new lot number must be entered, press

followed by

. A

screen for entry of the cuvette lot number is displayed:

Enter the new cuvette lot number, using the number keys, then press

. A confirmation screen is displayed.

10.

Press

followed by

to confirm the cuvette lot number. A screen for final

confirmation of the liquid control lot number and cuvette lot number is displayed:

followed by

11.

Press

to confirm the lot numbers. Test results are displayed

within ten seconds.

12.

Record the THb, %O

2

Hb, %COHb, and %MetHb results, or print the results of the

test (see page 35).

13.

Verify that the results are within the expected range of values for the liquid control.

14.

If desired, plot the liquid control results on daily logs such as the ones shown in
the Quality Control Logs section on page 57.