44

DAC-530E-EN-BH

7.2 Downloading the Internal Data Log

Regardless of whether a DDC is installed in the unit, select information is recorded internally within
the DDU-120 AED. The information recorded is limited to:

• The time the AED was turned on.

• Other data such as event milestones (motion detection, shock advice, shock delivery

information, etc).

• Eight seconds of ECG data immediately before a shock/no-shock decision, eight seconds

immediately after each shock, and all ECG data during the charging and waiting-to-shock

periods.

• Note: Audio data is not logged internally.

To download the internally logged information, perform the following procedure:

• Insert a blank DDC into the unit.

• Turn the unit on.

• Once the unit is on, turn it off in data download mode by pushing and holding the ON/OFF

button for at least five seconds.

• Allow the unit to write the contents of the internal log to the DDC by waiting for the unit to

turn off automatically.

The DDU-120 will write the contents of the internal log onto the DDC. This information can then
be reviewed using the DefibView software.