Defibtech DDU-120 Series User Manual

Page 59

Advertising
background image

51

DAC-530E-EN-BH

8.1.7 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions and Immunity

Electromagnetic conformity

The DDU-120 is intended for use in the electromagnetic environment specified below. The
customer or the user of the DDU-120 should assure that it is used in such an environment.

Electromagnetic emissions

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions

CISPR 11

Group 1

Class B

The DDU-120 uses RF energy only for its internal

function. Therefore, its RF emissions are very low

and are not likely to cause any interference in nearby

electronic equipment.

The DDU-120 is suitable for use in all establishments,

including domestic establishments and those directly

connected to the public low-voltage power supply

network that supplies buildings used for domestic

purposes.

Harmonic emissions

IEC 61000-3-2

Not applicable

Voltage fluctuations/flicker

emissions

IEC 61000-3-3

Not applicable

Electromagnetic immunity

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment -

guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

The DDU-120 uses interference

detected and motion detected

indicators to notify the user if

conditions are not ideal. No other

ESD requirements are necessary.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power line

supply lines

±1 kV for input/

output lines

Not applicable

Surge

IEC 61000-4-5

±1 kV line(s) to

line(s)

±2 kV line(s) to earth

Not applicable