5 serial port connection, 6 commissioning – Flintec FAA-27 User Manual

Page 11

Advertising
background image

FAA-27, Technical Manual, Rev. 1.1, June 2014

Page 10 of 21


You can also set the analog output type after power on as;

Turn the programming switch to the analog output type ( to V or I position )

Connect or disconnect TxD and RxD pins according the table above to set the

output range.

Turn the adjustment switch to the

position. After 2 seconds turn in back to

the

0“ - position ( if the instrument is powered on before ).


The analogue output type can be followed by the V /

I LED on front of the instrument

after running it ( refer to table above ) .

After set up the analog output of the instrument according to the table above go to
the next step.

4.5 Serial Port Connection



4.6 Commissioning

Double check the following issues before energizing the instrument.

Check Mechanical installation, grounding, load cell connection and power

supply connection.

Check the analogue output set up described in

chapter 4.4

.

The analogue output cabling should be done for the same analogue output

type.

The adjustment (Adj) rotary switch shall be at “0“ - position.

Advertising