Flintec FAA-27 User Manual

Page 16

Advertising
background image

FAA-27, Technical Manual, Rev. 1.1, June 2014

Page 15 of 21

2. eCal procedure steps:

Descriptions

Step 1 START eCal

Command:

@011000000001020001EB

Start the eCal process.

Response:

@011000000001EE

ACK

Step 2. Total Load cell Emax entry

Command:

@01100003000204000001F4F1

Total LC is 500kg.

Response:

@011000030002EA

ACK


Step 3. Average Load cell mV/V entry

Command:

@0110000500020400004E2F67

LC sensitivity: 2.0015

mV/V

.

Response:

@011000050002E8

ACK


Step 4. Estimated Dead load entry

Command:

@0110000700020400000000E2

Dead load is 0kg.

Response:

@011000070002E6

ACK

Step 5. Scale Capacity entry

Command:

@01100009000204000000FAE6

Scale capacity is 250kg.

Response:

@011000090002E4

ACK


Step 6. SAVE with estimated dead load value

Command:

@011000000001020007E5

Save the eCal values.

Response:

@011000000001EE

ACK


Or SAVE with automatic zero adjustment.

Unload the scale and send the command below
to start the zero adjustment.
Command:

@011000000001020015D7

Save eCal with Zero Adj

Response:

@011000000001EE

ACK


Step 7. APPLY

Command:

@011000000001020009E3

Apply the calibration.

Response:

@011000000001EE

ACK

Step 8. STOP eCal

Command:

@011000000001020002EA

Stop the eCal process

Response:

@011000000001EE

ACK

3. Digital Filter Setting:

Command:

@011000220001020000CA

Set the Fast filter

Command:

@011000220001020001C9

Set to Medium filter

Command:

@011000220001020002C8

Set to Slow filter

Response:

@011000220001CC

ACK