Flintec FAA-27 User Manual

Page 13

Advertising
background image

FAA-27, Technical Manual, Rev. 1.1, June 2014

Page 12 of 21

Use the formula for calculating the analog output value

Analog

Output

= Minimum output +

Maximum output

– Minimum output

* Load

Scale capacity



Example:
Scale capacity

100kg

Output range

4

– 20 mA

Load

25kg


The 4-20mA analog output current will be at 25kg ;

I

out

= 4mA + (( 20mA

– 4mA ) / 100kg ) * 25kg = 4 + 0.16 * 25 =

I

out

= 8mA

The 0

– 10VDC analog output voltage will be at 25kg:


V

out

= 0 + (10V / 100kg ) * 25kg = 0,1 * 25 =

V

out

= 2.5VDC

Turn the adjustment switch to the G position from the analog output type

position you set before.

Increase or decrease the analog output by adjustment rotary switch to the

calculated output value.

Never bring the adjustment switch position to

position in this adjustment.

The adjustment switch must be at 0 - position at the end of the adjustment.

Turn the programming switch to analog output type position for operation

(to V or I).

Important note:

The instrument saves 4 different adjustments in its memory for the

voltage and current outputs and their ranges. Changing the analog output type it
automatically set the adjustment to the output range used before.

Advertising