Labconco PuriCare Bedding Disposal Stations 3840020 User Manual

Page 25

Advertising
background image

21

Product Service 1-800-522-7658

Chapter 4: Theory Of Operation And Safety Precautions


without first disconnecting the
PuriCare from electrical service.

Avoid the use of flammable gases or
solvents in the PuriCare. Care must
be taken to ensure against the
concentration of flammable or
explosive gases or vapors. An open
flame should NOT be used in the
PuriCare. Open flames may disrupt
the airflow patterns in the station or
damage the filters’ adhesive. Gases
under high pressure should not be
used in the PuriCare station, as they
may disrupt the airflow patterns of
the station.

The surface of the HEPA filters are
fragile and should not be touched.
Care must be taken to avoid
puncturing the HEPA filter during
installation or normal operation. If
you suspect that a HEPA filter has
been damaged, DO NOT use the
station; contact a local certification
agency or Labconco at 800-522-7658
or 816-333-8811 for recertification
information.

THE HEPA FILTER IN THE
PURICARE STATION WILL
GRADUALLY ACCUMULATE
AIRBORNE PARTICULATE
MATTER FROM THE ROOM AND
FROM WORK PERFORMED IN THE
STATION. THE RATE OF
ACCUMULATION WILL DEPEND
UPON THE CLEANLINESS OF THE
ROOM AIR, THE AMOUNT OF TIME
THE STATION IS OPERATING AND
THE NATURE OF WORK BEING
DONE IN THE STATION.

)

!