Chapter 6: Maintaining Your PuriCare
Prefilter replacement:
1. Turn the station off.
2. Loosen the four thumbscrews that secure the
prefilter access panel to the top of the work area, as
shown in Figure 6-1. Let the access panel swing
down.
3. Pull the prefilter straight out and down.
4. Install a new prefilter, orienting the prefilter airflow
directional arrows up towards the HEPA filter.
Postfilter replacement:
1. Remove the postfilter from its holder on top of the
station by pulling it forward, as shown in Figure
6-2.
2. Discard the old postfilter.
3. Remove the clear plastic wrapping off of the new
postfilter.