Starting the Station


To start the PuriCare, turn the blower switch to the
“ON” position, as shown in Figure 5-1.

Reading the Filter System
Indicator


The two light system located on the control panel as
shown in Figure 5-1 indicates the total system pressure.
During operation, a green “NORMAL” indicator light
stays on. When the pre- or HEPA filters become
excessively loaded, the “NORMAL” light turns off and
the “SERVICE” light turns on, indicating the unit
should be serviced. If the “SERVICE” light stays on
after replacing the prefilter, a qualified certifier can
then determine if the HEPA filter needs to be replaced.