Product Service 1-800-522-7658

Appendix C: Bedding Disposal Station Specifications

Environmental Conditions

• Indoor use only.

• Maximum altitude: 6562 feet (2000 meters).

• Ambient temperature range: 41° to 104°F (5° to

40°C).

• Maximum relative humidity: 80% for

temperatures up to 88°F (31°C), decreasing
linearly to 50% relative humidity at 104°F
(40°C).

• Main supply voltage fluctuations not to exceed

±10% of the nominal voltage.

• Transient overvoltages according to Installation

Categories II (Overvoltage Categories per IEC
1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for
115V models and 2500V for 230V models are
allowed.

• Used in an environment of Pollution degrees 2

(i.e., where normally only non-conductive
atmospheres are present). Occasionally,
however, a temporary conductivity caused by
condensation must be expected, in accordance
with IEC 664.


Advertising