33

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare


5. If the ‘NORMAL’ light is on, remove the dress

panel, and using the appropriate hex wrench,
turn the adjustment switch slowly clockwise as
shown in Figure 6-5. Reinstall the dress panel.
Repeat this procedure until the ‘SERVICE’ light
turns on when the dress panel is replaced.

6. If the ‘SERVICE’ light is on, remove the dress

panel, and using the appropriate hex wrench,
turn the adjustment switch slowly
counterclockwise as shown in Figure 6-5.
Reinstall the dress panel. Repeat this procedure
until the ‘NORMAL’ light turns on when the
dress panel is replaced.

7. To confirm correct operation, turn the unit off,

uncover the postfilter and restart the station.
The ‘NORMAL’ light will be on. When the
postfilter is slowly covered over, the
‘SERVICE’ light will illuminate.



HEPA Filter Replacement

The HEPA filter should be serviced
only by a qualified certifier. After the
HEPA filter is replaced, the unit
MUST be certified.

1. Unplug the cabinet.
2. Remove the front dress panel as described in

Customer Service Operations section of this
manual.

!

Figure 6-5