Requirements, Current calibration, Requirements current calibration – B&K Precision 9171-9174 - Manual User Manual

Page 171

Advertising
background image

159

Requirements

Before calibrating, the

“+”output terminal must be connected to the

“+S” terminal and the “–”output terminal must be connected to the

“–S” terminal using the included shorting bars/pins.

The following equipment is required for the calibration procedure:

6 ½ digit precision digital multimeter

Shunt for current calibration (Necessary if the maximum

output current of the power supply is greater than the

maximum current the multimeter is capable of measuring)

DC source with an output up to 10VDC and accuracy of at

least ±0.1% (for external analog input calibration)

5 kΩ resistor with accuracy within ±0.1% or better (for

external analog input calibration)

Current Calibration

- If the maximum output current of the power supply is within the

current limit of the DMM, then use the wire connections shown in

Figure 31 to calibrate current.

- If the maximum output current of the power supply is greater

than the current limit of the DMM, then a current shunt must be

used. The wire connections between the power supply and the

DMM will be setup like Figure 33.