Teledyne LeCroy WaveRunner 6 Zi and 12-Bit HRO Getting Started Manual User Manual
Page 168

WaveRunner 6 Zi and 12-Bit HRO
160
922136-00 Rev A
2. On the Delay Setting constant in area of the dialog, touch either the
Time or Div button.
As you change the timebase, this control allows you to either keep
the horizontal offset indicator stationary (when Div is selected) or to
have it move with the trigger point (when Time is selected). The
advantage of selecting Div is that the trigger point remains on the
grid as you increase the timebase; whereas, if Time is selected, the
trigger point could move off the grid.
Note
: Regardless of whether you select
Time
or
Div
, the
Delay
shown in
the timebase setup dialog always indicates time. However, when
Div
is
selected for Delay In, the delay in time is scaled proportional to the change
in timebase, thereby keeping the division on the grid constant.
3. If this checkbox control is available on your oscilloscope, you can
mark the Reset trigger counter before starting a new acquisition
checkbox on the Trigger Counter Setting area of the dialog.
This preference clears the trigger counter each time the oscilloscope
issues an arm acquisition command and applies when you have set a
trigger Holdoff condition in the Trigger dialog in either time or
events:
The default condition of this control is off (unchecked).
A
CQUISITION
S
TATUS
For each general category of oscilloscope operation, you can view a
summary of your setups. Access these status dialogs by making the
following menu bar selections:
Vertical → Channels Status
Timebase → Acquisition Status