LumaSense Technologies INNOVA 1412i User Manual

Page 65

Advertising
background image

Chapter 5

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 65 of 202

5.3.4

Full reset and/or Partial Reset of the Monitor via its

Front Panel

There are two well-defined levels of reset. By pressing RESET the

following text appears on the display screen:

If S3 is pressed, you can come out of reset mode without performing

any kind of reset. The Monitor just goes back to operating the same

way it was operating before the RESET push-button was pressed.

S1 PARTIAL Reset: If this level of reset is chosen the Monitor

stops operating, performs a partial reset as described in the

previous section (

section 5.3.2

) and then resumes operating in

the fashion described in

Table 5.1

.

S2 FULL Reset: If this level of reset is chosen the Monitor responds

by changing the text on the display so that you have a chance to

confirm that you wish the Monitor to perform a FULL reset:

S1 By pressing this push-button the user can come out of reset

mode without performing any kind of reset. The Monitor just

goes back to operating the same way it was operating before

RESET was pressed.

S3 By pressing this push-button the user confirms that a FULL

reset is required. During a FULL reset the Monitor performs the

following two tasks:

·

“Clearing” (emptying) all data from its Working Memory.

This means all data in Display Memory and Background

Memory will be lost.

SELECT RESET TYPE

PARTIEL FULL NONE

PERFORM FULL RESET ?

NO YES

Advertising