Harmonic analysis, Setting for interval, Thd calculation setting – KYORITSU 6310 Instruction Manual User Manual
Page 71
KEW6310 4.2.2 Measurement Setting (Harmonic Analysis)
KEW6310
4.30
Harmonic Analysis
Press the
F3
Key at each Measurement setting screen, and move to the screen for Harmonic Analysis Setting.
Setting for interval
* Default value (or after system reset) : 30 min
* Setting procedure is same to that for interval Setting for W/ Wh/ DEMAND.Refer to the procedure described in
the preceding pages.
Interval listed on the drop down list depends on the number of save items with “ON” setting. Alter the
number of save items with “ON” setting when desirable interval isn’t listed on the drop down list.
Interval of 1 sec is not available.
Interval
Number of “ON”
2 sec
1
5 sec
2
10 sec
5
THD Calculation Setting
THD stands for “Total Harmonic Distortion”.
THD-F THD-R
* Default value (or after system reset) : THD-F
THD-F Fundamental
waveform-basis
THD-R
Total RMS value-basis
1
Press the
Cursor
Keys and select [THD Calculation], and then press the
ENTER
Key.
2
Press the
Cursor
Keys and select “THD-F” or “THD-R”, and then press the
ENTER
Key.
Drop down list appears.
Selected THD calculation
method is displayed.