Harmonic analysis, Setting for interval, Thd calculation setting – KYORITSU 6310 Instruction Manual User Manual

Page 71

Advertising
background image

KEW6310 4.2.2 Measurement Setting (Harmonic Analysis)

KEW6310

4.30


Harmonic Analysis

Press the

F3

Key at each Measurement setting screen, and move to the screen for Harmonic Analysis Setting.

Setting for interval


* Default value (or after system reset) : 30 min

* Setting procedure is same to that for interval Setting for W/ Wh/ DEMAND.Refer to the procedure described in

the preceding pages.


Interval listed on the drop down list depends on the number of save items with “ON” setting. Alter the
number of save items with “ON” setting when desirable interval isn’t listed on the drop down list.
Interval of 1 sec is not available.

Interval

Number of “ON”

2 sec

1

5 sec

2

10 sec

5

THD Calculation Setting

THD stands for “Total Harmonic Distortion”.

THD-F THD-R

* Default value (or after system reset) : THD-F

THD-F Fundamental

waveform-basis

THD-R

Total RMS value-basis

1

Press the

Cursor

Keys and select [THD Calculation], and then press the

ENTER

Key.


2

Press the

Cursor

Keys and select “THD-F” or “THD-R”, and then press the

ENTER

Key.

Drop down list appears.

Selected THD calculation

method is displayed.

Advertising