4.2.2 Measurement Setting (QUALITY-Transient Measurement) KEW6310
4.49
KEW6310
Setting for threshold
Voltage Range
150V
300V
600V
1000V
Threshold
(on 1V basis)
50~310Vpeak 90~630Vpeak
170~1270Vpeak
340~2000Vpeak
* Default value (or after system reset) : 1415V
* Vrms value (Vpeak divided by √2) is automatically calculated when Vpeak is set.
1
Press the
Cursor
Keys and select [Threshold Value], and then press the
ENTER
Key.
2
Press the
Cursor
Keys and alter the values, and press the
ENTER
Key
to fix it.
Box with
▲▼
mark appears
at the rightmost digit.
Selected threshold value is
displayed. Value displayed in
parenthesis is a threshold value
divided by √2.