Laser emission stop input, Gain input – KEYENCE IG Series User Manual
Page 96

4-2 Basic Settings and Advanced Settings
4-38
4
Setting V
a
ri
o
us
F
u
nct
ion
s
IG-E
Laser emission stop input
The laser emission continues to stop while the laser emission stop input is ON. The laser
emission indicator on the sensor amplifier lights off while the laser emission is stopped.
"Amplifier control unit" (page 1-10)
Timing chart
*1 After 2.5 ms (max.), the internal measurement value (R.V.) becomes [
].
Minimum Input Time
ON time: 2 ms
OFF time: 2 ms
Gain input
When the Gain input is set to ON, the standard waveform is registered (Gain adjustment).
"3-4 Registering the Standard Waveform (Gain adjustment)" (page 3-9)
Timing chart
*1 1.5 sec. is taken when "Nonvolatile memory (EEPROM) storage function" is set for
"12. Saving the standard waveform function".
Minimum Input Time
ON time: 20 ms
OFF time: 100 ms
*1
*1 1.5 sec. is taken when "Nonvolatile memory (EEPROM) storage function" is set for
"12. Saving the standard waveform function".
Reference
While the laser emission is stopped, the internal measurement value (R.V.)
shows [
].
"2 Display Screen and Output" (page A-9)
Max. 2 ms*
1
Max. 2 ms
Min. 2 ms
ON
OFF
Laser emission
stop input
Laser being transmitted
Laser being transmitted
Laser emission being stopped
Max. 120 ms*
1
Min. 20 ms
Before registering the standard waveform
Registering
The standard waveform registration complete
ON
OFF
Gain input