:SOURce:HARMonic Commands

[:SOURce[<n>]]:HARMonic:AMPL

Syntax

[:SOURce[<n>]]:HARMonic:AMPL <sn>,{<value>|MINimum|MAXimum}

[:SOURce[<n>]]:HARMonic:AMPL? <sn>[,MINimum|MAXimum]

Description

Set the amplitude of the specified order of harmonic in the harmonic function of the

specified channel.
Query the amplitude of the specified order of harmonic in the harmonic function of

the specified channel.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

<sn>

Integer

2 to 8

2

<value>

Real

0Vpp to the amplitude upper limit of

the specified channel

1.264 7Vpp

Explanation

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

 The amplitude upper limit of the specified channel is limited by the

“Impedance” (

:OUTPut[<n>]:IMPedance

or

:OUTPut[<n>]:LOAD

) and

“frequency/period” (

[:SOURce[<n>]]:FREQuency[:FIXed]

or

[:SOURce[<n>]]:PERiod[:FIXed]

) settings.

Return Format

The query returns the harmonic amplitude in scientific notation with 7 effective

digits, for example, 1.000000E+00 (the harmonic amplitude is 1Vpp).

Example

:SOUR1:HARM:AMPL 5,1 /*Set the amplitude of the fifth order of harmonic of

CH1 to 1Vpp*/

:SOUR1:HARM:AMPL? 5 /*Query the amplitude of the fifth order of harmonic

of CH1 and the query returns 1.000000E+00*/

Related

Commands

:OUTPut[<n>]:IMPedance

:OUTPut[<n>]:LOAD
[:SOURce[<n>]]:FREQuency[:FIXed]
[:SOURce[<n>]]:PERiod[:FIXed]