Source[<n>]]:apply:triangle – RIGOL DG1000Z Series User Manual

Page 92

Advertising
background image

RIGOL

Chapter 2 Command System

2-80

DG1000Z Programming Guide

[:SOURce[<n>]]:APPLy:TRIangle

Syntax

[:SOURce[<n>]]:APPLy:TRIangle

[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum

[,<offset>|DEFault|MINimum|MAXimum[,<phase>|DEFault|MINimum|MAXimum]]]]

Description

Set the waveform of the specified channel to triangle waveform with the specified

frequency, amplitude, offset and phase.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

<freq>

Real

1uHz to 1MHz

1kHz

<amp>

Real

Refer to the “Explanation”

5Vpp

<offset>

Real

Refer to the “Explanation”

0V

DC

<phase>

Real

0° to 360°

Explanation

 The triangle waveform is ramp waveform with 100% symmetry.

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

 The range of <amp> is limited by the “Impedance” (

:OUTPut[<n>]:IMPedance

or

:OUTPut[<n>]:LOAD

) and “Frequency/Period” settings. The range of

<offset> is limited by the “Impedance” and “Amplitude/High Level” settings.

Example

:SOUR1:APPL:TRI 100,1,2,3 /*Set the waveform of CH1 to triangle waveform

with 100Hz frequency, 1Vpp amplitude, 2V

DC

offset

and 3° start phase*/

Related

Commands

:OUTPut[<n>]:IMPedance
:OUTPut[<n>]:LOAD