[:SOURce[<n>]]:APPLy:HARMonic

Syntax

[:SOURce[<n>]]:APPLy:HARMonic

[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum

[,<offset>|DEFault|MINimum|MAXimum[,<phase>|DEFault|MINimum|MAXimum]

]]]

Description

Enable the harmonic function of the specified channel and set the fundamental

waveform (sine) parameters (frequency, amplitude, offset and phase).

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

<freq>

Real

1uHz to 20MHz

1kHz

<amp>

Real

Refer to the “Explanation”

5Vpp

<offset>

Real

Refer to the “Explanation”

0V

DC

<phase>

Real

0° to 360°

Explanation

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

 The range of <amp> is limited by the “Impedance”

(

:OUTPut[<n>]:IMPedance

or

:OUTPut[<n>]:LOAD

) and “Frequency/Period”

settings. The range of <offset> is limited by the “Impedance” and

“Amplitude/High Level” settings.

 When this command is executed, the instrument uses the default harmonic

parameters or the harmonic parameters set at the last time. You can send the

[:SOURce[<n>]]:HARMonic

series commands to set the desired harmonic

parameters and enable or disable the harmonic function.

Example

:SOUR1:APPL:HARM 100,1,2,3 /*Enable the harmonic function of CH1 and set

the fundamental waveform (sine) parameters

to 100Hz frequency, 1Vpp amplitude, 2V

DC

offset and 3° start phase*/

Related

Commands

:OUTPut[<n>]:IMPedance
:OUTPut[<n>]:LOAD
[:SOURce[<n>]]:HARMonic