Tp velocity at loss of material, Hhour meter current value, Icurrent & max temperature – Laser beta lasermike LS8000-3 User Manual

Page 134

Advertising
background image

LaserSpeed 8000-3 Instruction Handbook

RS-232/RS-422 Communication

Part No. 93463 / Drawing No. 0921-01561

Page 134 of 221

Revision A (Sep 2007)

TP

Velocity at Loss of Material

TP<CR>

Read

Response

Format: <N>

TP<N><CR>

Set

<N>

is an integer either 0 or 1 indicating the Velocity at Loss of Material setting.

This setting can be used to force the measured Velocity and QF to 0 if Material Present is in
the "Material Not Present" state. This can be used, for example, to force the pulse outputs to
stop counting in some situations.

By default, the LS8000 Velocity and Quality Factor outputs are not affected by Material
Present.

Value

Description

0

(default)

Hold Last Reading

Velocity and Quality Factor are not affected by Material
Present state. The gauge will continue to measure if
possible, and will output the measured velocity. If
measurements are lost, and the last velocity was above
the Hold Velocity if Above setting, then the velocity output
will be held at this value.

1 Zero


Velocity and Quality Factory are both forced to zero when
in the "Material Not Present" state, regardless of whether
measurements are being made.

H

Hour Meter Current Value

H<CR>

Read Only Command

Response

Format:

<N>

This command returns the total number hours that the gauge has been running.

I

Current & Max Temperature

I<CR>

Read Only Command

Response

Format: <N>,<N>

This command returns both the current internal temperature, and the maximum temperature
that has been recorded. Both temperatures are in Celsius. The values returned from the
LS8000-3 must be divided by 100 to get the temperature.

The normal temperature rise from ambient to internal is 10-15ºC.

Example:

Command: I<CR>