Ethernet – Laser beta lasermike LS8000-3 User Manual

Page 218

Advertising
background image

LaserSpeed 8000-3 Instruction Handbook

Appendix E: Troubleshooting Guide

Part No. 93463 / Drawing No. 0921-01561

Page 218 of 221

Revision A (Sep 2007)




Shutter Open




Material Present















System Ready

connected to User Ground (Pin
11,19,21) for the laser to turn on. Check
wiring connections.

The Laser Interlock (Pin 17) must be
connected to User Ground (Pin
11,19,21) for the laser to turn on. Check
wiring connections.

Check setting in LaserTrak for
Internal/External Material Present

Internal: Adjust Material Present

threshold level until Material Present
bit changes.

External Active Low: Disconnect

Material Present (Pin 10), or connect
to User Ground to turn Material
Present On, and Connect to +5-
24V

DC

to turn Material Present Off.

External Active High: Disconnect

Material Present (Pin 10) to turn OFF
Material Presnent, and connect to +5-
24V

DC

to turn ON Material Present.

Make sure you reference the voltage
input to the User Ground (Pin
11,19,21).

System Ready bit is a combination of
the Laser On, Shutter Open, and Laser
At Temp status bits. These three bits
must all be ON for the System Ready bit
to be on.

Nothing illuminated on gauge (i.e.
no LED lights, and laser is off)

Loss of Power




Power Supply Failure

The POWER ON Led should always be
illuminated when the gauge has power.
Check electrical connections, fuses, and
power switch

1. Check power supply voltage
2. Replace Power Supply

Ethernet


Can't get Ethernet to work.

No Connection




Gauge doesn't have

Check the Link status light on the router
to see if the connection is ok. Set the
gauge's COMM LED to indicate Ethernet
traffic.

Send the gauge the *ETHERNET?<CR>