PN 89-00002-00-072 Rev. C

6 - 1

Maintenance and Cleaning

To ensure accurate test results, properly clean and maintain the
Luminex 100 IS system. Read and follow all instructions in this
section. For your convenience, a maintenance log form is included at
the end of this chapter.

Do not remove the analyzer cover under any circumstances!

Daily Maintenance

If the system is powered on, but idle for more than four hours, click
the Maintenance tab. Click Warmup. Wait 30 minutes for the
Luminex 100 analyzer and the optics system to warm up.

Before Running Samples

‹

Before you run samples:

1. Turn the Luminex 100 analyzer on. The system begins warming

automatically.

2. Verify the levels of sheath fluid and waste fluid.

3. When the system is warmed up, click Prime to prime the

analyzer, then click OK.

4. Click Alcohol Flush. A confirmation dialog box opens. Click

Eject/Retract to eject the Luminex XYP instrument tray. Place
at least 1.2 mL of 70% isopropanol or 70% ethanol in the
reservoir. Click OK.

6

Warning:

When analyzing potentially infectious biological

samples on the Luminex 100 analyzer, follow standard
laboratory safety practices. These safety precautions should also
be taken when cleaning or maintaining the analyzer.