Ns. see figure 5-4 – Luminex 100 IS Version 2.3 User Manual

Page 48

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 6

PN 89-00002-00-072 Rev. C

1. Microtiter plate/reservoir image 2. Temperature and Pressure Gauges

3. Command List

Figure 5-4. Run Batch Tab

The command buttons on the Run Batch tab are:

For more information on these commands, see the Commands
section, starting on page 5-31. For more information on setting up
batches, see See “Batch Setup Procedures” on page 51..

The microtiter plate and reservoir image represents where you
place samples or other fluids used in running or maintaining the
system. Samples are analyzed vertically, from top to bottom within
the column, and then from left to right for subsequent columns.

The Temperature and Pressure Gauges show information about
the analyzer and the XYP instrument. The XYP heater temperature
measures the internal Luminex XYP instrument plate temperature.

1

3

2

Print Worklist

Eject/Retract

Start Plate

Resume

Cancel Command

Pause

Cancel all

Advertising