Flush the system, Monthly, Clean the sample probe – Luminex 100 IS Version 2.3 User Manual

Page 144: Clean exterior surfaces, Flush the system -4, Monthly -4

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

6 - 4

PN 89-00002-00-072 Rev. C

Flush the System

Run 3 backflushes, 3 drains, 2 alcohol flushes and 3 washes with
distilled water.

Monthly

Clean the Sample Probe

‹

To clean the sample probe:

1. Remove the sample probe as follows. First, unsnap the light

housing located above the probe. Then, unscrew the Cheminert
fitting on top of the probe completely. Next, gently grasp the
probe and push up. Remove the probe out of the top of the
sample arm.

2. Clean the sample probe using a bath sonicator or using a 10 mL

syringe. If you are using a bath sonicator, place the smaller end
of the sample probe in the bath sonicator for 2 to 5 minutes. If
you are using a syringe, force 10% to 20% bleach through the
larger end of the sample probe.

3. Replace the sample probe and adjust the vertical height. You

should adjust the vertical height anytime the probe is removed.

4. Alcohol Flush the system with 70% isopropanol or 70% ethanol.

Clean Exterior Surfaces

‹

To clean exterior surfaces, follow these steps:

1. Disconnect the system from ac power by turning off the power

switches and unplugging the Luminex 100 analyzer, the
Luminex XYP instrument, and the Luminex SD system.

2. Wipe all exterior surfaces with mild detergent, then a 10% to

20% bleach solution, and finally with plain distilled water.

Danger:

Be sure that the system is not performing an operation

when you remove the sample probe.

Caution:

The Luminex XYP instrument sample probe should slide

up easily while removing it from the sample arm. If you feel
resistance, do not force the probe up. Contact Luminex Technical
Support.