Luminex 100 IS Version 2.3 User Manual

Page 92

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 50

PN 89-00002-00-072 Rev. C

4. Ensure that the analyzer is set to draw CON1 and CON2 beads

from the wells you loaded in step 2 of “Run System xMAP
Calibrators” on page 5-47
.

5. In the Maintenance tab, click CON1 then click OK. Wait until

CON1 completes. The Device Status section in the status bar
changes from “Running” to “Standby”.

6. Click CON2, then click OK. Wait until CON2 completes. The

Device Status section in the status bar changes from “Running”
to “Standby”. The System Monitor on the Diagnostics tab
displays date and time in green if both CON1 and CON2 are
successful.

7. Ensure the analyzer is set to the draw distilled water from the

well you loaded it in step 2 of “Run System xMAP Calibrators”
on page 5-47
.

8. Click Wash to wash the system after running the system

calibrators. Wash a total of four times. You will need to change
the well location. Click on the dropdown menu located to the
right of the Wash button.

9. Click OK and wait until the wash completes. The device status

section in the status bar changes from “Running” to “Standby”.

If an error occurs during system calibration or verification, an X
appears in front of the Diagnostics tab title, and the title text turns
red.

Selecting Existing

CAL or CON Lots

‹

To select an existing lot:

1. On the Maintenance tab, click New CAL Targ or New CON

Targ.

2. Select a lot using the arrows located to the right of the Import

and Export buttons in the Update CAL Targets and Update CON
Targets dialog boxes.

Review the lot information and press OK to select the calibration lot.

Importing CAL or

CON Lots

‹

To import system CAL or CON lots:

1. On the Maintenance tab, click New CAL Targets or New CON

Targets as appropriate. An Update CAL Targets or Update
CON Targets
dialog box opens.