To reprocess samples using Replay Batch:

1. On the Acq. Detail tab, click Replay Batch. The Browse for

Folder dialog box opens displaying the My Batches folder.

2. Select the desired batch under the My Batches folder and click

OK.

3. The Open Template dialog box opens. Click on the desired

template and click Select.

4. The Run Batch tab becomes the active tab. You can monitor the

commands as they process. Click on the Acq. Detail tab and
monitor the data, histogram, and dot plot.

After replaying a batch, you can analyze the data.

‹

To analyze Replay Batch data:

1. Click Start Analysis.

2. In the Open Batch dialog box, select the batch you want to

analyze, then click Select. The most recent Replay Batch is the
last or has the highest ID number. See Figure 5-29.

.

Figure 5-29. Analyze Open Batch

The Analysis window opens, showing the Batch info in the
Standards tab. To close the Analysis window click Close. See
“Analyzing Batches and Multi-Batches” on page 76..

Note:

The Open Batch

dialog box does not list or
show the templates asso-
ciated to the batches.