Calibrating the Probes

3.2

CALIBRATING THE LEVEL TROLL INSTRUMENTS

The Level TROLL instrument has been factory-calibrated with NIST standards to a
greater degree of accuracy than can be achieved in nearly any alternate setting.

The Level TROLL has the ability to be “zeroed” in air, but should never be “zeroed”
in water. Overall, In-Situ Inc. does not recommend zeroing the sensor.

3.2.1

Zero Pressure (Air Only)

1. From the Con TROLL PRO Main menu, select Calibrate, Level TROLL, then

Zero Pressure.

3.2.2

Restoring Defaults

If you have zeroed the Level TROLL and wish to return to the factory settings:

1. From the Con TROLL PRO Main menu, select Calibrate, Level TROLL, then

Defaults.

3.3

CALIBRATING THE AQUA TROLL INSTRUMENTS

From the Con TROLL PRO Main menu, select Calibrate, then Aqua TROLL 200 (or
100). You will have the following options:

1 Conductivity

2 Zero Pressure

3 Defaults

3.3.1

Calibrating Conductivity

The Aqua TROLL 200 instrument’s factory calibration with NIST-traceable
standards guarantees a high degree of linearity across the entire operating range of
5 to 100,000 µS/cm. This is described as a “flat” response. The Aqua TROLL 200 is
capable of meeting its published specifications without requiring additional
calibration by the user, as most commercially available standards can introduce a
larger potential measurement error than the instrument’s initial factory calibration.

A user calibration is only recommended if you must conform to a standard
operating procedure or if the cell has undergone physical change (e.g., deposits on
cell walls that cannot be removed; physical damage to cell walls).

1. If you opt to calibrate the Aqua TROLL conductivity cell, refer to the calibration

procedure in the Aqua TROLL Operator’s Manual for laboratory setup
instructions.

2. From the Con TROLL PRO Main menu, select Calibrate, then Aqua TROLL 200

(or 100), then Conductivity.

3. Use the arrow keys to enter the standard value. Select Enter.
4. You will see the Calibration Beginning icon next to the µS/cm value. When the

calibration reaches stability, choose Stable to accept the value or Cancel to
return to the pre-existing calibration (Figure 7).

FIGURE 7. Beginning the calibration; accepting or cancelling the calibration

Cancel

Conductivity

12102 μS/cm

23.1 °C

Stable

Cancel

Conductivity

12050 μS/cm

23.1 °C

Calibration beginning

Accept or cancel calibration

Nominal

Cancel

Conductivity

12050 μS/cm

23.1 °C

Accept or cancel calibration