Outputs and Relays

4.2.1

Notes About Alarms, Setpoints, and Deadband Values

In high alarm mode:

• The relay is turned on if the current reading is equal to or greater than the

setpoint value.

• The relay is turned off if the current reading is less than or equal to the setpoint

minus the deadband (Figure 8).

FIGURE 8. High alarm relay behavior

In low alarm mode:

• The relay is turned on if the current reading is equal to or less than the setpoint

value.

• The relay is turned off if the current reading is greater than or equal to the

setpoint plus the deadband (Figure 9).

FIGURE 9. Low alarm relay behavior

4.3

HOLD (PAUSE) THE OUTPUTS

Occasionally, you may wish to temporarily pause the outputs.

1. Select the Main menu, Hold Outputs. All measurement updates will stop and the

current values and outputs will be held until you select Resume.

2. You can use the View function and arrow keys to scroll through the parameters

on Line 1 and Line 2.

Time

Parameter value

Relay ON

Relay OFF

High Alarm

Deadband

Setpoint

Time

Parameter value

Relay ON

Relay OFF

Low Alarm

Deadband

Setpoint