Connect the pvc blower drain, Connect to the blower outlet, For coated steel blowers – Labconco Coated Steel, Fiberglass and PVC Blowers 7183200 User Manual

Page 15

Advertising
background image

11

Product Service 1-800-522-7658, International 816-333-8811

Chapter 3: Getting Started

Connect the PVC Blower Drain


The PVC Blower has a 1/2" NPT drain connection in the bottom of
the housing. When this connection is used, it should be directed
into a proper drain or into the exhaust duct for proper disposal.

CAUTION: Draining the blower housing
directly onto the roof may cause damage to your
roof due to the corrosive chemicals exhausted.

Connect to the Blower Outlet


Before proceeding with the blower outlet connection, read the two
warnings listed below:

WARNING: Should your exhaust stack, on the
outlet side of the exhaust blower, extend over 7
feet, both guy wires and additional structural
supports are required to carry the weight of this
ductwork. The guy wires must be substantial
enough to support the exhaust stack against high
wind velocities.

WARNING: If the blower includes an integral
backdraft damper, do not interfere with or limit
the travel of the damper mechanism when
connecting exhaust ductwork.

For Coated Steel Blowers


Coated Steel Blowers model numbers 7068000 through 7068700,
include a 10" by 5-1/2" rectangular outlet. A rectangular to round
Transition Adaptor, Labconco part number 4722401, adapts the
outlet connection to accept 10-inch diameter PVC ductwork
Transition Adaptor. Labconco part number 4722400 adapts the
outlet on these blowers to accept 8-inch diameter PVC ductwork.

Coated Steel Blowers, model numbers 7068800 through 7069700
include a 13-1/2" by 7" rectangular outlet. Transition Adaptor,
Labconco part number 7003400, adapts the outlet of these blowers
to accept 12-inch diameter PVC ductwork.

PVC

!

!

CS

Advertising
This manual is related to the following products:

Coated Steel, Fiberglass and PVC Blowers 7183100, Coated Steel, Fiberglass and PVC Blowers 7183000, Coated Steel, Fiberglass and PVC Blowers 7182300, Coated Steel, Fiberglass and PVC Blowers 7182200, Coated Steel, Fiberglass and PVC Blowers 7182100, Coated Steel, Fiberglass and PVC Blowers 7182000, Coated Steel, Fiberglass and PVC Blowers 7181700, Coated Steel, Fiberglass and PVC Blowers 7181600, Coated Steel, Fiberglass and PVC Blowers 7181500, Coated Steel, Fiberglass and PVC Blowers 7181400, Coated Steel, Fiberglass and PVC Blowers 7181300, Coated Steel, Fiberglass and PVC Blowers 7181200, Coated Steel, Fiberglass and PVC Blowers 7181100, Coated Steel, Fiberglass and PVC Blowers 7181000, Coated Steel, Fiberglass and PVC Blowers 7180900, Coated Steel, Fiberglass and PVC Blowers 7180800, Coated Steel, Fiberglass and PVC Blowers 7180700, Coated Steel, Fiberglass and PVC Blowers 7180600, Coated Steel, Fiberglass and PVC Blowers 7180500, Coated Steel, Fiberglass and PVC Blowers 7180400, Coated Steel, Fiberglass and PVC Blowers 7180300, Coated Steel, Fiberglass and PVC Blowers 7180200, Coated Steel, Fiberglass and PVC Blowers 7180100, Coated Steel, Fiberglass and PVC Blowers 7180000, Coated Steel, Fiberglass and PVC Blowers 7071900, Coated Steel, Fiberglass and PVC Blowers 7069700, Coated Steel, Fiberglass and PVC Blowers 7069600, Coated Steel, Fiberglass and PVC Blowers 7069500, Coated Steel, Fiberglass and PVC Blowers 7069400, Coated Steel, Fiberglass and PVC Blowers 7069300, Coated Steel, Fiberglass and PVC Blowers 7069200, Coated Steel, Fiberglass and PVC Blowers 7069100, Coated Steel, Fiberglass and PVC Blowers 7069000, Coated Steel, Fiberglass and PVC Blowers 7068900, Coated Steel, Fiberglass and PVC Blowers 7068800, Coated Steel, Fiberglass and PVC Blowers 7068700, Coated Steel, Fiberglass and PVC Blowers 7068600, Coated Steel, Fiberglass and PVC Blowers 7068500, Coated Steel, Fiberglass and PVC Blowers 7068400, Coated Steel, Fiberglass and PVC Blowers 7068300, Coated Steel, Fiberglass and PVC Blowers 7068200, Coated Steel, Fiberglass and PVC Blowers 7068100, Coated Steel, Fiberglass and PVC Blowers 7068000