Chapter 7: modifying your blower, Thermoplastic duct, Duct couplings, female – Labconco Coated Steel, Fiberglass and PVC Blowers 7183200 User Manual

Page 41

Advertising
background image

37

Product Service 1-800-522-7658, International 816-333-8811

Chapter 7: Modifying Your Blower

Nominal Diameter/Inches

6

8

10

12

16

Actual OD/inches

6.625

8.625

10.750

12.750

16.000

Actual ID/inches

6.25

8.250

10.375

12.375

15.625

Catalog Number

47086

47189

70272

56020

56050

Shipping Wt./lbs.

20

35

50

65

80

Airflow/CFM

Static Pressure Loss/Inches H

2

0

For Each 10 ft. of Duct Length

250

.039

.011

.003

.001

--

500

.147

.037

.013

.005

.001

750

.321

.079

.026

.011

.003

1000

.557

.140

.043

.018

.005

1250

.855

.210

.066

.027

.008

1500

--

.300

.095

.039

.012

1750

--

.380

.130

.053

.016

2000

--

.485

.155

.067

.020

2500

--

--

.245

.109

.031

3000

--

--

--

.145

.042

4000

--

--

--

.240

.074

5000

--

--

--

--

.120


Thermoplastic Duct

PVC exhaust duct is Type 1, unplasticized, schedule 40,
lightweight and corrosion-resistant. A female duct coupling is
required to join two sections. Connections are simple with solvent
cement. This rigid duct may be cut without special tools. Comes
in 10' lengths.

Nominal Diameter/Inches

6

8

10

12

16

Catalog Number

47086

47189

70272

56020

56050

Actual OD/inches

6.625

8.625

10.750

12.750

16.000

Actual ID/inches

6.250

8.250

10.375

12.375

15.625

Shipping Wt./lbs.

25

35

50

65

80

Duct Couplings, Female

PVC coupling makes connection between two sections of
thermoplastic duct quick and easy.

Nominal Diameter/Inches

6

8

10

12

16

Catalog Number

47089

47192

70275

56023

56053

Shipping Wt./lbs.

4

5

5

6

7

Equivalent Resistance in Feet
of Straight Duct

0

0

0

0

0

Advertising
This manual is related to the following products:

Coated Steel, Fiberglass and PVC Blowers 7183100, Coated Steel, Fiberglass and PVC Blowers 7183000, Coated Steel, Fiberglass and PVC Blowers 7182300, Coated Steel, Fiberglass and PVC Blowers 7182200, Coated Steel, Fiberglass and PVC Blowers 7182100, Coated Steel, Fiberglass and PVC Blowers 7182000, Coated Steel, Fiberglass and PVC Blowers 7181700, Coated Steel, Fiberglass and PVC Blowers 7181600, Coated Steel, Fiberglass and PVC Blowers 7181500, Coated Steel, Fiberglass and PVC Blowers 7181400, Coated Steel, Fiberglass and PVC Blowers 7181300, Coated Steel, Fiberglass and PVC Blowers 7181200, Coated Steel, Fiberglass and PVC Blowers 7181100, Coated Steel, Fiberglass and PVC Blowers 7181000, Coated Steel, Fiberglass and PVC Blowers 7180900, Coated Steel, Fiberglass and PVC Blowers 7180800, Coated Steel, Fiberglass and PVC Blowers 7180700, Coated Steel, Fiberglass and PVC Blowers 7180600, Coated Steel, Fiberglass and PVC Blowers 7180500, Coated Steel, Fiberglass and PVC Blowers 7180400, Coated Steel, Fiberglass and PVC Blowers 7180300, Coated Steel, Fiberglass and PVC Blowers 7180200, Coated Steel, Fiberglass and PVC Blowers 7180100, Coated Steel, Fiberglass and PVC Blowers 7180000, Coated Steel, Fiberglass and PVC Blowers 7071900, Coated Steel, Fiberglass and PVC Blowers 7069700, Coated Steel, Fiberglass and PVC Blowers 7069600, Coated Steel, Fiberglass and PVC Blowers 7069500, Coated Steel, Fiberglass and PVC Blowers 7069400, Coated Steel, Fiberglass and PVC Blowers 7069300, Coated Steel, Fiberglass and PVC Blowers 7069200, Coated Steel, Fiberglass and PVC Blowers 7069100, Coated Steel, Fiberglass and PVC Blowers 7069000, Coated Steel, Fiberglass and PVC Blowers 7068900, Coated Steel, Fiberglass and PVC Blowers 7068800, Coated Steel, Fiberglass and PVC Blowers 7068700, Coated Steel, Fiberglass and PVC Blowers 7068600, Coated Steel, Fiberglass and PVC Blowers 7068500, Coated Steel, Fiberglass and PVC Blowers 7068400, Coated Steel, Fiberglass and PVC Blowers 7068300, Coated Steel, Fiberglass and PVC Blowers 7068200, Coated Steel, Fiberglass and PVC Blowers 7068100, Coated Steel, Fiberglass and PVC Blowers 7068000