B. probe/sample handling, B. probe/sample handling -22, Model microcem – Emerson MicroCEM User Manual

Page 36

Advertising
background image

Instruction Manual

748467-A
January 2002

1-22 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Model MicroCEM

Paramagnetic

O

2

Electro-

Chemical

O

2

NDIR

CO

Chemiluminescense

NO

X

Linearity

<

± 1%

<

± 1%

<

± 1%

<

± 1%

1

Zero Drift

<

± 1% /day

<

± 1% /day

<

± 1% /day

<

± 1% /day

1

Span Drift

<

± 1% /day

<

± 1% /day

<

± 1% /day

<

± 1% /day

1

Repeatability

<

± 1%

<

± 1%

<

± 1%

<

± 1%/day

1

Response Time (t90)

10<

± t90< ±-15

10<

± t90< ± 15 15s< ± t90< ± 20s 15s< ±-t90< ± 20s

Influence of Ambient

Temperature

(-20

°C to 45°C)

On Zero
On Span

<

± 1%

<

± 1%

<

± 1%

<

± 1%

<

±-2%

<

±-2%

<

±-2%

<

±-2%

b. Probe/Sample Handling

Power ............................................ Universal Power Supply 85 – 264 VAC, 50 – 60 Hz, + 10%

500 Watts Maximum at Start Up. 250 Watts Nominal

Mounting........................................ Wall Mount
Area Classification......................... General Purpose / NEMA 4X (IP65) Fiberglass Enclosure
Ambient Range

Temperature .......................... -30

° to 50° Celsius

Relative Humidity................... 5 to 99%

Instrument Weight ......................... 75 lbs. Typical
Dimensions.................................... 24“ x 24“ x 12“ (HxWxD)
Stack Sample Moisture ................. Up to 25%
Sample Cooler............................... Thermo Electric dual pass Chiller.

Permeation Tube (-30

° C) dewpoint.

Customer instrument air required @ 5 L/M, -40

° C dewpoint

Max. Stack Temperature ............... 500

° F (Higher temperatures available by utilizing elongated

spools)

Stack Pressure .............................. -5 to 15 inches H

2

O

Sample Flow Rate ......................... 1 L/min from sample handling enclosure to Analysis enclosure
Response Time (Max distance between Analysis Enclosure and Sample Conditioning/Probe)

Enclosure is 300'. (Response time is 20 seconds/100' w/1/4" tub-

ing).

Probe Length................................. 48" length 316 SS Probe with .5 micron sintered filter. (Customer

to cut to length in field.)

Mounting Flange............................ Optional 4“ 150#
Sample Pump................................ 316 SS diaphragm type
Instrument Air Requirements......... Instrument grade air required. 15 SCFM @ 60 -100 PSIG (30 sec-

onds 2 times per day). (Pressure Regulation by Customer.)

1

0-10ppm NOx range is <

± 3%.