d. Manual Calibration

A dry-run Calibration may be initiated
from the Manual Calibration page of the
MicroCEM Settings. The results of the
calibration will not be applied and only
provide a dry run of the calibration. If
desired a partial calibration may be in-
voked for one or more of the emission
types. While the manual calibration is
in process, a calibration progress dialog
will be displayed as shown in

Figure

3-26. When the manual calibration is
completed, the results are displayed in
the Manual Calibration Results dialog
as shown in Figure 3-12. If the Local
Calibration checkbox is checked, the
Local Calibration valve will be used
during the calibration rather than the
probe Calibration valve.

NOTE

“Start Autocalibration now” will in-
voke a real calibration and will apply
new correction factor results when
done.