0 maintenance and service, 1 overview, Maintenance and service -1 – Emerson MicroCEM User Manual

Page 83: Overview -1, Model microcem

Advertising
background image

Instruction Manual

748467-A

January 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 5-1

Model MicroCEM

SECTION 5

MAINTENANCE AND SERVICE

DANGER

.

ELECTRICAL SHOCK HAZARD

Disconnect power to the module(s) prior to
replacing components.

WARNING

QUALIFIED PERSONNEL

This equipment should not be adjusted or
repaired by anyone except properly quali-
fied service personnel.

WARNING

PARTS INTEGRITY

Tampering with or unauthorized substitu-
tion of components may adversely affect
safety of this product. Use only factory-
approved components for repair.

5-1 OVERVIEW

The MicroCEM Analyzer Module requires very
little maintenance during normal operation.

Occasionally, the detector's reaction chamber
and sapphire window may require cleaning,
refer to Section 5-5.

White crystal deposits on the windows of the
reaction chamber and plugging of capillaries
and vent are usually due to sample contami-
nates such as ammonia reacting with the high
ozone levels and NO components. To elimi-
nate the contaminates, the sampling system
should be reworked or a preventive mainte-
nance program developed (if dropout is not
excessive). Another source of crystalline for-
mation is contaminated air.

Several components may require replace-
ment. These are discussed in the following
sections.