2 converter, 2 converter assembly, Converter -3 – Emerson MicroCEM User Manual

Page 85: Figure 5-2, Converter assembly -3, Model microcem, Instruction manual

Advertising
background image

Instruction Manual

748467-A

January 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 5-3

Model MicroCEM

5-2 CONVERTER

To replace the converter or sensor, discon-
nect the two pneumatic tubes and two electri-
cal connections. Unlace the heater blanket,

and remove the converter. Reassemble in re-
verse order, ensuring that the converter is ori-
ented with the glass cloth at the bottom and
the sensor is oriented correctly inside the
heater jacket.

Figure 5-2. Converter Assembly

ASSEMBLED SIDE VIEW

Converter

Tube 655227

Wrap with

aluminum foil

Glass

Cloth

Sensor

Heater

Jacket

655228

Sensor

655282

Advertising