Detector type: fid-mc – execution screen – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 287
![background image](/manuals/360216/287/background.png)
<5. EtherLCD>
5-36
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
FID-MC
Det operation
○
0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
Reset status
○
0: Initialized
1: Setting done
If the voltage value is zero,
Initialized is displayed;
otherwise Setting done.
Detector type: FID-MC – Execution screen
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Det Adjustment
Oven #
1 Det # 1 ( FID-MC )
>
- Det operation Off
MC voltage value 0.0 V
- Target MC current
Main
<<
<
>
>>
F1
F2
F3
F4
F5
F6
Figure 5.47
Example of the detector type: FID-MC – execution screen
F2 (Main):
Displays the Det Adjustment (Reset) screen.
F3 (<<):
Decreases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).
F4 (<):
Decreases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).
F5 (>):
Increases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).
F6 (>>):
Increases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).
• To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent
key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the
bottom.
• If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255
and the command is successfully sent internally, “Setting succeeded” is displayed. Users
can confi rm the results with the voltage value and current value refreshed every second.
• Selecting the Target MC voltage displays Target MC voltage (AA.A/B.B):. If the value
17.5/0.5 is entered, it will be shown as 17.5+/–0.5 V, 5.0 as 5.0+/– V and 2/1 as 2.0+/– 1.0 V.
• The numbers of ovens to which no detectors are installed cannot be set via F5.