5 method setup, Gcm method setup screen (main) – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 301
![background image](/manuals/360216/301/background.png)
<5. EtherLCD>
5-50
IM 11B08A01-01E
5.4.5 Method
Setup
GCM method setup screen (Main)
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
GCM Method Setup
GCM #1 / Method #1
>
- Main cycle 60.0 s
- Warming up time 5.0 s
- Stream valve ON time
-5.0 s
- Stream valve OFF time 0.0 s
- Pause time 55.0 s
Main
Seq
Sys
Copy
GCM# Method#
F1
F2
F3
F4
F5
F6
Figure 5.63
Example of the GCM method setup screen
F1(Menu):
Displays the Table Menu screen.
F2(Seq):
Displays the GCM Sequential screen.
F3(SYS):
Displays the SYS Method screen.
F4(Copy):
Displays the GCM Method Copy screen.
F5(GCM #): Specifi es the GCM number.
F6(Method #): Specifi es the Method number.
• Pressing F3 (SYS) displays the SYS Method Setup screen with the smallest number among
the SYS numbers set to GCM.
• Pressing F5 (GCM #) displays GCM number: in the second line on the bottom. Entering the
method number refl ects the GCM number in the second line on the top and displays the
GCM method setup assigned to that number.
• Pressing F6 (Method #) displays Method number: in the second line on the bottom. Entering
the method number refl ects the method number in the second line on the top and displays
the GCM method setup assigned to that number.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
GCM #
GCM number
1
6
Method #
Method number
1
6
Main cycle
○
10
See right
seconds Upper Limit will depend on
sample rate
When 20 ms: 2700 seconds
When 40 ms: 5400 seconds
When 80 ms: 10800 seconds
When 160 ms: 21600 seconds
* The smallest sample rate set
for the detector used by the SYS
belonging to the relevant GCM.
Warming up
time
○
5.0
9999.9
seconds
Stream valve
ON time
○
−Warming
up time
0.0
seconds
Stream valve
OFF time
○
0.0
Main cycle
− 2
seconds
Pause time
○
5
Main cycle
− 2
seconds
2nd Edition : May 11, 2012-00