4 detector setup – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 299
![background image](/manuals/360216/299/background.png)
<5. EtherLCD>
5-48
IM 11B08A01-01E
5.4.4 Detector
Setup
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Detector Setup
(1/2)
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Detector Setup
(2/2)
Oven #
1 Det # 1 ( TCD )
Oven #
1 Det # 1 ( TCD )
>
- Filtering const Off
>
- Auto gain setting None
- Sample rate 20 ms
- Full scale value 0
- Square root calc req * * * * *
- Auto zero Off
- Flame detect level * * * * *
- Auto zero value 0.0000 mV
- Sense set Equal
- Auto zero time 0.0 s
- Input resistor * * * * * * * * * *
Menu
Oven#
Det#
Menu
Oven#
Det#
F1
F2
F3
F4
F5
F6
F1
F2
F3
F4
F5
F6
Figure 5.62
Example of the detector setup screen
F1 (Menu):
Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #):
Sets a detector number.
TIP
• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *
(asterisk).
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
None
TCD
FID
FID-MC
FPD
Filtering const
○
0.001
1.000
Sample rate
○ *
20 ms
40 ms
80 ms
160 ms
Square root
calc req
○
The value is shown except
when the FPD is installed.
No
Yes
* The square root of the area
value
Flame detect
level
○
–20
20
mV
The item is displayed only when
the FID, FPD or FID-MC is
installed.
Sense set
○
No amplifi cation
Tenfold
Input resistor
○
1 Mohm
6 Mohm
20 Mohm
60 Mohm
200 Mohm
600 Mohm
1 Gohm
2 Gohm
*:
The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00