Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 295
![background image](/manuals/360216/295/background.png)
<5. EtherLCD>
5-44
IM 11B08A01-01E
• The initial screen displays the smallest oven number among those assigned to isothermal
ovens on the Temperature Control Set.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Vent #
Vent number
1
6
Carrier gas
number
Carrier gas number
---
1-1
1-2
2-1
2-2
3-1
3-2
* --- indicates not used.
The value is set on the Detector
Set screen.
Vent type
Vent type
○
---
BF
FF
Split
* --- indicates not used.
Vent fl ow rate Vent fl ow rate
○
Alpanumeric: 10 characters
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Operating Parameters
(5/8)
>
- Pressure unit kPa
- Vent fl owrate unit ml/min
- Sample volume unit ml
Menu
F1
F2
F3
F4
F5
F6
Figure 5.57
Example of the operating parameters screen (5/8)
F1 (Menu):
Displays the Table Menu screen.
TIP
If the pressure unit is changed, the gas pressure value including the one set to EPC controlled will be converted
according to the new unit. For example, if the unit is changed from kPa (the value range is set to 0 to 500 kPa) to
psi (the value range 0 to 72.5 psi), 250 kPa is converted to and displayed as 36.25 psi.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Pressure unit
○
kPa
psi (see the defi nition of SI
units.)
Vent fl ow rate
unit
○
ml/min
Sample
volume unit
○
ml
μl
2nd Edition : May 11, 2012-00