<5. EtherLCD>
5-69
IM 11B08A01-01E
Table 5.3
Synchronization rules table
Destination
Source
Measurement Cal
Val
Calibration
run
Sample run
Measurement
○
Ч
○
Ч
Ч
Cal
▲
Ч
▲
Ч
Ч
Val
○
Ч
○
Ч
Ч
Calibration run
Ч
Ч
Ч
Ч
Ч
Sample run
Ч
Ч
Ч
Ч
○
○, ▲: Synchronization possible, see Synchronize items for synchronization items.
*:
The
▲ symbol in the Synchronization Rules Table signifi es (Cal →Measurement, Cal →Val) and he synchronization items ▲
(process assignment, reference peak, etc.) cannot be synchronized.
×:
Cannot be synchronized.
The following is an example of setting error.
Synchronized destination: Stream 2ーPeak 1
Stream 1ーPeak 2
F0502.ai
Stream 1ーPeak 1
Indirect synchronization
Contradicts the rule “Cannot set peak within the same stream to synchronized destination.”