Epc program set screen – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 306
![background image](/manuals/360216/306/background.png)
<5. EtherLCD>
5-55
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Copy source
GCM #
GCM number
1
6
SYS #
SYS number
1
6
Method #
Method number
1
6
Copy target
GCM number
○
1
6
Copy target
method
number
○
1
6
EPC Program Set Screen
This screen is displayed when EPC has been set.
Press F4 (EPC).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
EPC program Set
(1/2)
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
EPC program Set
(2/2)
GCM #1 / SYS #1 / Method #1 / Gas #1-#1
GCM #1 / SYS #1 / Method #1 / Gas #1-#1
>
- Initial Press setpoint ZZZ ZN NAAA
>
- #1 Press ramp rate
ZZN NAAA/min
- #1 Press ramp start time ZZZ ZN Ns
- #1 Press hold setpoint ZZZ ZN NAAA
Total run-time ZZZ ZN Ns
- #1 Press hold time ZZZ ZN Ns
- #2 Press ramp rate ZZN NAAA/min
- #2 Press hold setpoint ZZZ ZN NAAA
- #2 Press hold time ZZZ ZN Ns
- #3 Press ramp rate ZZN NAAA/min
- #3 Press hold setpoint ZZZ ZN NAAA
- #3 Press hold time ZZZ ZN Ns
Main
Oven#
Gas#
Main
Oven#
Gas#
F1
F2
F3
F4
F5
F6
F1
F2
F3
F4
F5
F6
Figure 5.69
Example of the EPC program set screen
F1(Main):
Navigates to SYS Method Setup screen (Main).
F5(Oven#): Specifi es an oven number.
F6(Gas#): Specifi es a carrier gas number
• Only the gas belonging to the relevant SYS can be set with F6 (Gas#).
2nd Edition : May 11, 2012-00